Catalog Number 326638 |
Device Problems
Loose or Intermittent Connection (1371); Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/22/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the stopper inside the bd ultra-fine¿ ii insulin syringe was loose and misaligned on the plunger.The following information was provided by the initial reporter, translated from japanese: "before use of product, customer noticed that the black gasket inside the syringe was wobbly and tilted, making it difficult to measure accurately.".
|
|
Manufacturer Narrative
|
Medical device expiration date: unknown.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
|
|
Event Description
|
It was reported that the stopper inside the bd ultra-fine¿ ii insulin syringe was loose and misaligned on the plunger.The following information was provided by the initial reporter, translated from japanese: "before use of product, customer noticed that the black gasket inside the syringe was wobbly and tilted, making it difficult to measure accurately.".
|
|
Manufacturer Narrative
|
H.6.Investigation summary: samples were received and an investigation was performed.Embecta was able to duplicate or confirm the indicated issue.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
|
|