Model Number DB-1200-S
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/19/2023
Event Type  Injury  
Event Description
It was reported that during the second stage of the deep brain stimulation (dbs) implant procedure where the lead and lead extension were being implanted, it was found that the already implanted implantable pulse generator (ipg) set screw was unable to be tightened in order to secure the lead.The physician used the torque screwdriver to attempt to tighten the set screw however, he was unable to hear the clicking sound that indicates the torque limit.Therefore, the ipg was explanted and replaced intraoperatively.The explanted ipg will not be returned as it was retained by the medical facility.There were no resulting patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERCISE GEVIA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key17951122
MDR Text Key325897887
Report Number3006630150-2023-06314
Device Sequence Number1
Product Code NHL
UDI-Device Identifier08714729984443
UDI-Public08714729984443
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/21/2021
Device Model NumberDB-1200-S
Device Catalogue NumberDB-1200-S
Device Lot Number739589
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/19/2023
Initial Date FDA Received10/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/22/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age72 YR
Patient SexMale
Patient Weight106 KG
Patient RaceWhite
-
-