The user facility has informed richard wolf gmbh of an issue regarding a flex.Biopsy forceps 7fr wl 375mm, part id: 829.07, serial number # 4500377830.According to the received information, in (b)(6) 2023, difficulties occurred with the forceps during several operations.During a dj ureteral stent procedure, the pe (biopsy) forceps made it difficult to remove the samples because the instrument did not opened and closed properly.
|
The flex.Biopsy forceps 7fr wl 375mm, part id: 829.07, batch 4500377830 was manufactured on 24/oct/2022.The batch 4500377830 consisted of 43 pieces.The examination of the customer's forceps has shown that the instructions for use (ga-s004 / en / us /v3.0 / 2021-11 / pk21-0310 (rw: en / eu / v3.0 / 2021-11/ pk21-0310) regarding instrument reprocessing has not been fully followed.In chapter 10.1 reprocessing procedure - flexible forceps, the following is described under inspection and maintenance on page 21: "care after machine/manual disinfection: lightly oil the jaw section at the joint using 1-2 drops of instrument oil.Apply 2-3 drops of oil to the cleaning connection and open and close the jaw section.Ensure that all other surfaces are free of oil.Wipe away any excess oil." the jaw sections could be moved without any defects after oiling.The defect could have been detected early according to chapter 8 checks."injury due to damaged or incomplete products! injuries of the patient, user and others are possible.Run through the checks before and after each use.Do not use the products if they are damaged, incomplete or have loose parts.Do not attempt to do any repairs yourself." in chapter 8.2 function checks, the function is explicitly checked "check that the jaw sections open and close properly." in summary, a user error can be assumed.The incident could have been avoided if the information in the operating instructions has been observed.
|