MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSERT-IQ¿ ICM; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSERT-IQ¿ ICM; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION) Back to Search Results
Model Number DM5500
Device Problems Failure to Interrogate (1332); Pacemaker Found in Back-Up Mode (1440); Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2023
Event Type  malfunction  
Event Description
It was reported that patient presented for insertable cardiac monitor (icm) implant procedure.Post procedure it was noted that the icm was in back-up mode and was failing to interrogate.The icm was restored to normal device function.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of the device resetting was confirmed.Based on the information provided, it was determined that the reset reason was power-on reset (por).The root cause for the por was unable to be determined, but appeared to be likely be related to an external factor that occurred during implant procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSERT-IQ¿ ICM
Type of Device
RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION)
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17958314
MDR Text Key325958836
Report Number2017865-2023-48121
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067047809
UDI-Public05415067047809
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K230286
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDM5500
Device Lot NumberS000091619
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/29/2023
Initial Date FDA Received10/18/2023
Supplement Dates Manufacturer Received01/24/2024
03/20/2024
Supplement Dates FDA Received01/29/2024
03/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexFemale
Patient Weight97 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-