ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES
|
Back to Search Results |
|
Model Number 7122Q/65 |
Device Problems
Difficult to Fold, Unfold or Collapse (1254); Retraction Problem (1536); Capturing Problem (2891); Device Dislodged or Dislocated (2923)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/13/2023 |
Event Type
Injury
|
Event Description
|
Related manufacturer reference number: 2017865-2023-50085.It was reported a patient's right ventricular (rv) lead and atrial lead presented with high capture thresholds due to dislodgement, confirmed via x-ray.This was addressed in a procedure on (b)(6) 2023, in which both lead helixes would not spiral out normally, extend or retract.The rv and atrial leads were explanted and replaced within the same operation.Post-procedure the patient was stable.
|
|
Manufacturer Narrative
|
The reported events were lead dislodgement, helix mechanism issue and high pacing threshold.As received, a complete lead was returned in one piece with the helix partially extended clogged with blood/tissue.The reported event of helix mechanism issue was confirmed while the reported event of high pacing threshold was not confirmed.X-ray showed that the inner coil inside the connector region was over torqued consistent with procedural damage.After cleaning blood/tissue from the helix and applying torque to the connector pin, the helix could be fully extended and retracted.The full measured helix extension length was within specification.The cause of the reported event of helix mechanism issue was isolated to the helix being clogged with blood/tissue and over torqued of the inner coil which is consistent with procedural damage.Electrical tests and x-ray examination did not find any indication of conductor fractures or internal shorts.Visual inspection of the lead did not find any anomalies.
|
|
Search Alerts/Recalls
|
|
|