AIZU OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2023
Event Type  malfunction  
Event Description
An olympus onsite support specialist reported on behalf of the customer that the bending rubber damaged-metal protruding on the uretero-reno videoscope.The issue was found during preparation before use inspection for an unspecified therapeutic procedure.No patient was involved during the reported issue.The intended procedure was completed with another similar device without delay.Patient condition had no negative reports following the procedure, no documentation had been noted for any such occurrences.There were no reports of patient or user harm associated with this event.
 
Manufacturer Narrative
The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was returned to olympus for inspection, and the customer's complaint was confirmed.Device evaluation found scope bending section cover is damaged/leaking from cracked glue.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to the following: during device handling by the user, physical stress was applied to bending section, which led to occurrence of a-rubber (bending section cover) cut.However, metal did not protrude out of a-rubber.The event can be prevented by following the instructions for use (ifu) which state: "-inspection of the endoscope -do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17959336
MDR Text Key326634258
Report Number9610595-2023-15420
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170343582
UDI-Public04953170343582
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/22/2023
Initial Date FDA Received10/18/2023
Supplement Dates Manufacturer Received12/15/2023
Supplement Dates FDA Received12/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/21/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-