Intuitive surgical, inc.(isi) has requested that the prograsp forceps instrument be returned for evaluation.However, it has not yet been received as of the date of this report.A follow-up mdr will be submitted if the instrument is returned (post failure analysis evaluation) or if additional information is received.
|
It was reported that during a da vinci-assisted myomectomy surgical procedure, the prograsp forceps instrument broke and was unable to be removed through the cannula.The customer was able to remove the instrument with the cannula.They will return the instrument with the cannula as they cannot get the cannula off of the broken instrument.They believed they had retrieved all of the fragments from the instrument.Intuitive surgical, inc.(isi) followed up with the customer on and obtained additional information: the procedure was completed robotically.There was a delay of roughly 30 minutes.The instrument is available for return.The small fragments were discarded and would not be returned.The instrument was inspected prior to use with no noted damage.The reason the instrument and cannula were removed together was due to the wrist not being able to be straightened.The port size was increased to remove the cannula and instrument together.The incident did involve fragments falling inside the patient as a result of the instrument issue.The surgeon confirmed that all fragments were removed during the same procedure.There were no post-operative tests like an x-ray or ultrasound performed to check for remaining fragments.The patient did not return to the hospital due to experiencing any post-surgical complications related to retaining a foreign object.
|