ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number ASKU
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Intraocular Pressure Increased (1937); Nausea (1970); Pain (1994); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 09/01/2023
Event Type  Injury  
Event Description
A physician reported that during a cataract/vitrectomy combination surgery the surgeon has used anterior vitrectom with an ophthalmic operating console the used vitrectom turned out to be incompatible with the system.The issue caused delay in surgery necessitating its interruption after suturing the incisions.The patient was given with topical anesthesia and due to the delay in surgery and patient experienced ocular hypertonia , pain and nausea.Surgery was resumed under general anesthetic and was completed using another ophthalmic system and the patient condition is in stable now.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
Additional information provided in sections h.6.And h.10.Specific product identifiers (serial number) were not provided and could not be determined at this time.However, all device history records are reviewed prior to product release to ensure the product was manufactured in compliance with the device master record and meets release criteria.A review for complaints reported against this serial number cannot be performed as the serial number is unknown.The serial is unknown.Therefore, a service history review cannot be performed.Based on the information available, the customer reported event cannot be confirmed.Based on the information available, the customer reported event cannot be confirmed.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17963090
MDR Text Key326004950
Report Number2028159-2023-01416
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberASKU
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/10/2023
Initial Date FDA Received10/19/2023
Supplement Dates Manufacturer Received11/29/2023
Supplement Dates FDA Received12/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
Patient Age78 YR
Patient SexMale
-
-