Medtronic received information that prior to use of a bioconsole 560 instrument, it was reported a battery error and damaged user interface cable.The instrument was replaced.There was no patient involvement, so no adverse effect occurred.Medtronic received additional information that there were no power supply issues noted with the instrument.The battery has been installed in the instrument for 4 years, the battery met its life span.The lot number of the battery replaced is 9733878.
|
Device evaluation:the reported battery error and damaged user interface cable was verified during service.The issue was resolved by replacing the cable 560 user interface assy and the battery,12v,6.5 ah,certified.Preventive maintenance was performed per specifications.Note the device was not returned to medtronic facility bur was serviced by field service technician medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|