Catalog Number 03246353001 |
Device Problems
Degraded (1153); Mechanical Problem (1384); Material Integrity Problem (2978); Non Reproducible Results (4029)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/27/2023 |
Event Type
malfunction
|
Event Description
|
There was an allegation of analyzer alarms and questionable ise indirect gen 2 results from the cobas 6000 c 501 module.The initial chloride result was 6088.1 mmol/l and the repeat result was 100.1 mmol/l.The questionable result was not reported outside of the laboratory.
|
|
Manufacturer Narrative
|
The cobas 6000 c 501 module serial number was (b)(6).The investigation is ongoing.
|
|
Manufacturer Narrative
|
The field service engineer found there was a failing electrode.The customer replaced the na and k electrodes, internal standard (is), diluent, and reference.The ise was primed and activated and ise checks were performed that were successful.Calibration and qc were performed and were successful.The investigation determined the service actions resolved the issue.
|
|
Search Alerts/Recalls
|