• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION BIPAP AUTOSV; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION BIPAP AUTOSV; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Model Number DSX900H11C
Device Problem Degraded (1153)
Patient Problems Asthma (1726); Respiratory Tract Infection (2420); Lethargy (2560)
Event Date 10/22/2021
Event Type  Injury  
Event Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The patient has alleged sinus infection as well as upper respiratory infection and tiredness.Due to infection patient got notified by health provider to stop the use of the device.The patient alleged not having a working device has cause her asthma to get worse and is leaving her more tired.The device was returned to the manufacturer's product investigation laboratory for investigation.An external and internal visual inspection was completed by the manufacturer.The manufacturer found evidence of sound abatement foam degradation. the device's downloaded event log was reviewed by the manufacturer and found no error logged.The device provided airflow correctly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION BIPAP AUTOSV
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key17967942
MDR Text Key326079389
Report Number2518422-2023-26769
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090539
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX900H11C
Device Catalogue NumberDSX900H11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2023
Initial Date Manufacturer Received 10/22/2021
Initial Date FDA Received10/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1974-2021
Patient Sequence Number1
Patient Outcome(s) Other;
-
-