Model Number 3100A
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2023
Event Type  malfunction  
Event Description
It was reported to vyaire medical that the 3100a ventilator start/stop button is always on pressurizes as soon as the unit is powered on.The customer confirmed that there is no patient involvement associated with this reported event.
 
Manufacturer Narrative
Vyaire medical file identification: (b)(4).H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.H3 other text : device not returned yet.
 
Manufacturer Narrative
Device evaluation: g3, g6, h2, h3, h6 and h10.Result of investigation: vyaire medical arrived onsite to troubleshoot.Bypassed blender and humidifier.The customer using medical air tree set up instead of blender tee.As a resolution, replaced with correct set up.The seal on water trap was loose almost disconnected completely.Corrected and powered on unit.No alarms present at start up, unit passed both patient circuit check and performance check.Allowed the unit to ventilate with driver running, no issue.Reset/power and start/stop are both operating normally.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
sandra valencia
510 technology drive
irvine, CA 92618
8477865998
MDR Report Key17967959
MDR Text Key326079489
Report Number2021710-2023-18295
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/21/2023
Initial Date FDA Received10/19/2023
Supplement Dates Manufacturer Received11/06/2023
Supplement Dates FDA Received11/20/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-