Vyaire medical file identification: (b)(4).H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.H3 other text : device not returned yet.
|
Device evaluation: g3, g6, h2, h3, h6 and h10.Result of investigation: vyaire medical arrived onsite to troubleshoot.Bypassed blender and humidifier.The customer using medical air tree set up instead of blender tee.As a resolution, replaced with correct set up.The seal on water trap was loose almost disconnected completely.Corrected and powered on unit.No alarms present at start up, unit passed both patient circuit check and performance check.Allowed the unit to ventilate with driver running, no issue.Reset/power and start/stop are both operating normally.
|