• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION SUPERION INDIRECT DECOMPRESSION SYSTEM; PROSTHESIS, SPINOUS PROCESS SPACER/PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION SUPERION INDIRECT DECOMPRESSION SYSTEM; PROSTHESIS, SPINOUS PROCESS SPACER/PLATE Back to Search Results
Model Number 101-9810
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/26/2023
Event Type  malfunction  
Event Description
It was reported that during a superion indirect decompression system implant procedure while attempting to deploy the implant using the driver, the implant was catching on the lamina.The physician disengaged the implant to move it back just a little.Then tried to re-engage the implant multiple times but was unable to successfully.The patient started to decline, and the physician was not comfortable proceeding with the procedure.The procedure was rescheduled for a later date.
 
Event Description
It was reported that during a superion indirect decompression system implant procedure while attempting to deploy the implant using the driver, the implant was catching on the lamina.The physician disengaged the implant to move it back just a little.Then tried to re-engage the implant multiple times but was unable to successfully.The patient started to decline, and the physician was not comfortable proceeding with the procedure.The procedure was rescheduled for a later date.
 
Manufacturer Narrative
The returned spacer implant was analyzed, passed all tests performed, and exhibited normal device characteristics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERION INDIRECT DECOMPRESSION SYSTEM
Type of Device
PROSTHESIS, SPINOUS PROCESS SPACER/PLATE
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key17970793
MDR Text Key326321348
Report Number3006630150-2023-06388
Device Sequence Number1
Product Code NQO
UDI-Device Identifier00884662000536
UDI-Public00884662000536
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101-9810
Device Catalogue Number101-9810
Device Lot Number28501591
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/26/2023
Initial Date FDA Received10/19/2023
Supplement Dates Manufacturer Received12/01/2023
Supplement Dates FDA Received12/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
Patient Weight73 KG
-
-