• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2023
Event Type  malfunction  
Event Description
Philips received a complaint by the customer on the v60 indicating that there was an error for pressure sensor autozero failure.The device was in use on a patient at the time the reported issue was discovered; however, there was no harm to the patient or user.The device was swapped out with a different device.No medical intervention provided to the patient, nor a delay was noted.Per review of the diagnostic report, it was confirmed that the error for a pressure sensor autozero failure had occurred.Repair is currently pending.
 
Manufacturer Narrative
E1.(b)(6).
 
Manufacturer Narrative
A field service engineer (fse) went onsite and confirmed the reported issue via confirmation of error in the event log.The fse replaced solenoid 3 and 4 to resolve the reported issue.The device passed the required performance verification tests per philips standards and was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key17975783
MDR Text Key326160676
Report Number2518422-2023-27427
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838033832
UDI-Public00884838033832
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1076709
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/04/2023
Initial Date FDA Received10/20/2023
Supplement Dates Manufacturer Received11/01/2023
Supplement Dates FDA Received11/09/2023
Date Device Manufactured02/09/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-