ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Catalog Number 21AGFN-756
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6) 2023, a 21mm regent aortic mhv,flex cuff was selected for an implant.The physician rotated the valve with the valve rotator and the holder handle fully inserted into the orifice at a 90 degree angle.During the procedure, the physician discovered that one leaflet was detached and was removed from the patient.A 19mm regent aortic mhv,flex cuff was selected and implanted as a replacement.The patient is stable.
 
Manufacturer Narrative
An event of one leaflet detached during procedure was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Two images from field appeared to show a mechanical valve with one leaflet dislodged and placed on the side.The sewing cuff appeared red in color.Based on the information received, the potential cause of the reported incident could be related to oversizing, given that the procedure was successfully completed with a smaller valve.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Event Description
It was reported that on (b)(6) 2023, a 21mm regent aortic mhv,flex cuff was selected for an implant.The physician rotated the valve with the valve rotator and the holder handle fully inserted into the orifice at a 90 degree angle.During the procedure, the physician discovered that one leaflet was detached and was removed from the patient.A 19mm regent aortic mhv,flex cuff was selected and implanted as a replacement.The patient is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key17976640
MDR Text Key326166933
Report Number2135147-2023-04594
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005975
UDI-Public05414734005975
Combination Product (y/n)N
Reporter Country CodeLE
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number21AGFN-756
Device Lot NumberC00011760
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/28/2023
Initial Date FDA Received10/20/2023
Supplement Dates Manufacturer Received11/16/2023
Supplement Dates FDA Received12/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/05/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial