Model Number X SERIES |
Device Problem
No Display/Image (1183)
|
Patient Problem
Chest Pain (1776)
|
Event Date 09/29/2023 |
Event Type
malfunction
|
Event Description
|
Complainant alleged that while attempting to monitor a patient (age & gender unknown), the device's screen turned completely white and was illegible.Complainant indicated that the clinician obtained another device to continue treating the patient.Complaintant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed and attributed to user mis-handling.The technician observed damages to the device that were not consistent with normal wear and tear.The lcd display module was replaced.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
|