U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number X SERIES
Device Problem No Display/Image (1183)
Patient Problem Chest Pain (1776)
Event Date 09/29/2023
Event Type  malfunction  
Event Description
Complainant alleged that while attempting to monitor a patient (age & gender unknown), the device's screen turned completely white and was illegible.Complainant indicated that the clinician obtained another device to continue treating the patient.Complaintant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report was observed and attributed to user mis-handling.The technician observed damages to the device that were not consistent with normal wear and tear.The lcd display module was replaced.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key17979369
MDR Text Key326320111
Report Number1220908-2023-04001
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946005979
UDI-Public00847946005979
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/30/2023
Initial Date FDA Received10/20/2023
Supplement Dates Manufacturer Received09/30/2023
Supplement Dates FDA Received11/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-