• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG CUTTING LOOP, BIPOLAR, 24/26 FR.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG CUTTING LOOP, BIPOLAR, 24/26 FR. Back to Search Results
Model Number 27040GP1
Device Problem Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2020
Event Type  malfunction  
Event Description
It was reported that too much heat had been fired through the loop.The distal cutting loop was dislodged from its housing shaft.There was damage to the proximal connection prongs, it looks like there was burning around the yellow plastic coating.No patient harm or complication during procedure.
 
Manufacturer Narrative
The previous reportability decision for this event was reversed based on a retrospective review.The affected device has been requested for investigation by the manufacturer.Device was returned for investigation.Evaluation findings: the distal cutting loop is dislodged from its housing shaft.There is damage to the proximal connection prongs, it looks like there is burning around the yellow plastic coating.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUTTING LOOP, BIPOLAR, 24/26 FR.
Type of Device
CUTTING LOOP, BIPOLAR, 24/26 FR.
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
christiane klaiber
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
MDR Report Key17981156
MDR Text Key326368182
Report Number9610617-2023-00308
Device Sequence Number1
Product Code FAS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061541
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27040GP1
Device Catalogue Number27040GP1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2021
Initial Date Manufacturer Received 10/16/2023
Initial Date FDA Received10/20/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-