C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER Back to Search Results
Catalog Number 0165L14
Device Problem Gel Leak (1267)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the sterile package of the foley catheter was wet due to leakage of lubricant.The representative confirmed that the package was unopened.The middle of the paper surface of the sterilization bag was checked for the signs of moisture.They could not confirm any tears or leaks in the lubricant packaging but the whole place of the lubricant packaging was wet.Per follow up information received via ibc on 06oct2023, the customer clarified that they could not confirm if the other components of the tray was wet as the package was unopened.They found a mark on the middle of the paper label of the sterilization bag that appeared to have been wet.The lubricant packaging inside was not found to be torn or leaking.Since the package was unopened, they could not see any tears or leaks in the lubricant packaging inside, but the entire lubricant packaging was wet.
 
Manufacturer Narrative
The reported event was confirmed ¿ supplier related.The product had caused the reported failure.Sample was evaluated and observed no abnormality with the returned sample.However, observed traces of water marks on the pouch plastic of the catheter.The jelly packet was examined and suspected of micro leakage of jelly.Performed ftir testing and confirmed that the water marks were traces of leaked jelly.The reported event is confirmed as supplier related and documented.A potential root cause for this failure could be generated at supplier site or during transit to bard, (example: defective / torn / broken components).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "[warnings] 1.Method for use (1) do not inflate the balloon in the urethra.[the urethra may be injured.] (2) do not pull the catheter hard.[the bladder/urethra may be injured.] 2.Applicable patients (1) patients with delirium who might pull out catheter [when patient tugs at catheter unconsciously, the bladder and urethra may be damaged.] [contraindications] 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (4) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients (1) patients with known allergy to silver coated catheter [shape, configuration and principles] balloon catheter: silicone; silver coating this product is made with bacti-guard® silver alloy coating. available in sizes 12 to 24 every 2fr 1.Balloon catheter [intended use & effect- efficacy] this device is an indwelling catheter in the bladder for urinary drainage.The surface of catheter is coated first with a trace amount of metallic silver, and then with polyurethane onto that, to inhibit proliferation of bacteria that may adhere to the catheter.[directions for use] 1.Method of use the device is intended for single use only and is not reusable.(1) cleanse the area around the external urethral meatus with the cotton balls immersed in the antiseptics.(2) lubricate the distal end of the catheter with water-soluble lubricant.(3) insert catheter into the urethral meatus and advance it until the balloon enters the bladder and urine flows out through the catheter.Using a needle-less syringe, infuse the specified volume of sterile water into the inflation lumen to inflate the balloon.(4) pull catheter to seat the balloon at the level of the bladder neck and secure placement.(5) to deflate balloon and remove catheter, insert a luer tip (needle-less) syringe in the inflation valve to allow the water to drain spontaneously.After balloon deflation, withdraw the catheter while confirming that no resistance is encountered.2.Precautions for use (1) when resistance is encountered in inserting catheter, stop the procedure and remove the catheter.(2) when deflating balloon, do not aspirate with a syringe by hands.[the inflation lumen for balloon deflation may be occluded by negative pressure, and as a result the catheter cannot be removed.] (3) when inserting catheter with a stylet, confirm that the stylet has reached the tip of the catheter, and make sure to keep the stylet in place inside the catheter during insertion.(4) no substance except sterile water should be used to inflate the balloon.[if contrast medium is used, balloon may burst.If normal saline is used, crystallized salt may occlude the inflation lumen to prevent deflation of the balloon.If air is used, air may escape to cause inadvertent deflation of the balloon so that the catheter may come out prematurely.] (5) do not wipe catheter surface with organic solvents such as alcohol.(6) do not aspirate urine through drainage funnel wall.(7) since movement of the body, etc.May twist or bend catheter to cause occlusion, care should be taken to fix the catheter securely.(8) when urinary flow cannot be noted, confirm that the catheter is neither occluded nor broken.[precautions] 1.Precautions for use (exercise caution when using the device in the following patients) 1) exercise caution when using the device in patients with high urinary calcium levels as encrustation on the balloon surface, catheter occlusion or damage may occur.2.Important precautions 1) when catheter is inadvertently removed, inspect the balloon and shaft of catheter for rupture, defect, etc.Before inserting a new catheter.2) when any part of the balloon and/or the catheter is missing, consider removing them using a cystoscope.3) when it is difficult to remove catheter by deflating the balloon, take appropriate measures according to the section ¿troubleshooting¿. when it is difficult to remove catheter by deflating the balloon (expressed as ¿removal-difficult case¿ hereinafter), take appropriate measures according to the following procedures.The following two methods are available for removal-difficult cases.A.Non-rupture method (sterile water is withdrawn without bursting the balloon.) b.Balloon-rupture method with balloon-rupture method, fragments of the ruptured balloon may remain in the bladder.Therefore, try non-rupture method first.A.Non-rupture method 1) attach luer tip syringe to the inflation valve.Inject an additional amount of sterile water into the inflation lumen and pump the plunger.2) if situation wouldn't be improved with 1), sever the inflation funnel of valve.(fig.1) 3) if situation wouldn¿t be improved with 2), sever the catheter shaft while holding it with forceps so that the distal segment may not be drawn into the urethra.(fig.2) 4) if situation wouldn't be improved with 3), insert a needle into the inflation lumen and pump the plunger.(fig.3) 5) if situation wouldn't be improved with 4), insert a fine steel wire through the inflation lumen of catheter.(fig.4) b.Balloon-rupture method 1) inject 100-200ml/cc of saline solution warmed to body temperature into the bladder through the drainage lumen, and then inject a large amount of water into the balloon through the inflation lumen with a needle to induce rupture.After rupture of the balloon, irrigate the bladder.(fig.5) 2) if situation wouldn't be improved with 1), attempt following procedures.1) under the radioscopic observation, infuse a contrast medium into the bladder, and burst the balloon by suprapubic puncture of the bladder.(fig.6) 2) in male patients, burst the balloon by puncture with a needle from the perineal (or suprapubic) region or through the rectum under ultrasonographic guidance.(fig.7) 3) in female patients, burst the balloon by insertion of a needle along the urethra.(fig.8) 3.Malfunction and adverse events 1) malfunction - catheter kinking, damage, rupture - difficulty or failure to remove the device - occlusion of catheter inner lumens - encrustation - accidental removal of the device due to leakage of sterile water or balloon rupture - device damage due to inappropriate use 2) adverse events - urinary-tract infection - hemorrhage, hematuria - allergy reaction to the device - calculus formation - edema - pain - discomfort - injury of bladder or urethral - urethritis, urinary incontinence - retained balloon fragments [storage method and expiration date] 1.Storage store in a dry, cool place away from heat, moisture, and direct sunlight.2.Expiration date indicated on the direct package and the outer box." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that the sterile package of the foley catheter was wet due to leakage of lubricant.The representative confirmed that the package was unopened.The middle of the paper surface of the sterilization bag was checked for the signs of moisture.They could not confirm any tears or leaks in the lubricant packaging but the whole place of the lubricant packaging was wet.Per follow up information received via ibc on 06oct2023, the customer clarified that they could not confirm if the other components of the tray was wet as the package was unopened.They found a mark on the middle of the paper label of the sterilization bag that appeared to have been wet.The lubricant packaging inside was not found to be torn or leaking.Since the package was unopened, they could not see any tears or leaks in the lubricant packaging inside, but the entire lubricant packaging was wet.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17981537
MDR Text Key326266316
Report Number1018233-2023-07629
Device Sequence Number1
Product Code EZC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0165L14
Device Lot NumberMYHN0104
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/28/2023
Initial Date FDA Received10/20/2023
Supplement Dates Manufacturer Received11/24/2023
Supplement Dates FDA Received12/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-