C.R. BARD, INC. (BASD) -3006260740 POWERPORT IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 1708000 |
Device Problem
Device Markings/Labelling Problem (2911)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/02/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, a photo and images were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiry date: 05/2024).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
|
Event Description
|
It was reported that sometime post port placement, upon an x-ray examination, the port allegedly did not showed the port identifiers in the x-ray image.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.Three x-ray images were provided for review.The first and third appear to show an implanted port.The ct mark is not able to be identified.The second image is a port still in the packaging with ct mark clearly identifiable.Therefore the reported device marking issue can be confirmed.A definitive root cause could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiry date: 05/2024).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that sometime post port placement, upon an x-ray examination, the port allegedly did not showed the port identifiers in the x-ray image.There was no reported patient injury.
|
|