ST. JUDE MEDICAL, INC.(CRM-SYLMAR) OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR; Permanent defibrillator electrodes Back to Search Results
Model Number LDA210Q/58
Device Problem Retraction Problem (1536)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/02/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented during implant procedure.During procedure, it was noted that the helix of the right ventricular lead was extended and could not be retracted.The reported lead was not installed and the procedure was completed with an alternative lead on (b)(6) 2023.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of helix failure to retract was confirmed.As received, a complete lead was returned in one piece for analysis.Visual and x-ray examination found the helix was bent and damaged due to procedural damage.The cause of the reported event of helix mechanism issue was isolated to the bent and damaged helix.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
Permanent defibrillator electrodes
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17986047
MDR Text Key326282622
Report Number2017865-2023-50377
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734507325
UDI-Public05414734507325
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLDA210Q/58
Device Lot NumberA000144234
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/02/2023
Initial Date FDA Received10/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexMale
-
-