RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Device Sensing Problem (2917)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/02/2023
Event Type  malfunction  
Manufacturer Narrative
E1.Reporting institution phone number - (b)(6).Reporter phone number - (b)(6).
 
Event Description
Philips received a complaint from the customer reporting a patient disconnect condition that resulted in low tidal volume on the v60 ventilator.The device was in clinical use.The user replaced the patient nasal catheter and exchanged the v60 for another ventilator to continue therapy.The patient reported chest tightness at the time of the event.No other clinical information or medical intervention was reported.Additional information has been requested regarding the reported event.
 
Manufacturer Narrative
A philips authorized service provider (asp) confirmed with the customer there was no injury or other harm resulting from this event.The customer biomedical engineering department resolved this issue by replacing the gas delivery system (gds).The provision of additional details requested regarding the clinical impact and the correction were declined by the customer.The device was operational and returned to service after repairs were completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key17988257
MDR Text Key326309097
Report Number2518422-2023-27608
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public00884838025776
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1076716
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/16/2023
Initial Date FDA Received10/23/2023
Supplement Dates Manufacturer Received10/24/2023
Supplement Dates FDA Received10/26/2023
Date Device Manufactured04/05/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-