Catalog Number 06481647190 |
Device Problems
Mechanical Problem (1384); High Test Results (2457)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/28/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The mg2 reagent lot number was 69798601 with an expiration date of 30-sep-2024.Before the event, calibration and qc were performed and they were not acceptable.After calibration and qc were repeated and were successful, the patient's sample was then tested.The field service engineer (fse) checked the cuvette usage records and the wash rinse unit and found no abnormalities.He observed white powder with worn cuvettes on the upper and lower parts of the cover plate.The fse cleaned the white powder and adjusted the fluid volume of the rinse station.The white powder was most likely the abrasion of the cell material by the metal cover.The root cause was due to the presence of the white powder.The investigation is ongoing.
|
|
Event Description
|
We received an allegation about discrepant results for 1 patient's sample tested with magnesium gen.2 (mg2) assay on a cobas 6000 c501 analyzer.Initial result: 2.01 mmol/l.No questionable result was released outside of the laboratory as the result did not match the patient's clinical diagnosis.And the sample was repeated.Rerun result: 0.83 mmol/l.
|
|
Manufacturer Narrative
|
The field service engineer checked the analyzer and adjusted the cuvette rinse station.He also cleaned the black water tank.The root cause was consistent with a maintenance issue.The service actions (adjusting the cuvette rinse station and cleaning the black water tank) resolved the issue.No further issues were reported afterward.
|
|
Search Alerts/Recalls
|