AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-S190-10
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found that due to damage on the video plug and a cut on charged coupled device cable, the color tone of the image was abnormal.In addition to the reportable malfunction, the following were noted: due to a scratch on the electrical contact of the video connector and damage on the charged coupled device unit, a noisy image occurred; due to wear of the angle wire, the bending angle in the up direction did not meet the standard value; the adhesive on the bending section cover had a chip; switch 1 had discoloration.Additionally, the following components had a scratch: the upward/downward angulation plate, the video connector case, the video connector, the angulation lever, the control unit, the grip, the light guide connector, the light guide-cover glass, the right/left lever, the right/left plate, switch 1, and switch 2.A review of the device history record found the device had a non-conformity during manufacture, however the device was shipped in accordance with specifications, and therefore, had no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, it is likely that one of the following led to the malfunction: breakage to the image sensor unit, such as disconnection caused by stress of repeated use, external factors, handling, or a defect on the components mounted on the electric circuit board, including the integrated circuit chip and capacitor.However, a definitive root cause cannot be identified.This issue is addressed in the instructions for use (ifu): ¿chapter 3 preparation and inspection, 3.8 inspection of the endoscopic system" confirm that the wli and nbi endoscopic images are normal.1.Before inspection, wipe the objective lens using clean lint-free cloths moistened with saline solution or sterilized water.2.Observe the palm of your hand in the wli and nbi endoscopic images.3.Confirm that light is output from the endoscope¿s distal end.(see figure 3.23) 4.Adjust the brightness level as appropriate.5.Confirm that the wli and nbi endoscopic images are free from noise, blur, fog, or other irregularities.6.Turn the angulation control levers slowly in each direction until it stops.7.Confirm that the wli and nbi endoscopic images do not momentarily disappear or display any other irregularities." olympus will continue to monitor the field performance of this device.
 
Event Description
The customer reported to olympus that the endoeye flex deflectable videoscope had image noise.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was found: damage on the video plug causing an abnormal color tone of the image.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17990460
MDR Text Key326389536
Report Number9610595-2023-15691
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170310386
UDI-Public04953170310386
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/27/2023
Initial Date FDA Received10/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-