Model Number ACT100
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/27/2023
Event Type  malfunction  
Manufacturer Narrative
Device evaluation:the reported issue that the instrument was not passing wet quality controls was not verified during service.Every time the service technician ran electronic quality controls, it passed.However, the service technician spoke to technical support and proceeded to replace the harness assy actii , the actuator harness assembly , and the roller act ii.A second simulation of electronic quality controls 100, 200, and 500 was performed successfully multiple times.Preventive maintenance was performed per specifications.Note the instrument was not returned to medtronic facility but was serviced by field service technician medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that prior to use of an act plus instrument, it was reported that the instrument was not passing the wet quality controls.The instrument was replaced.There was no patient involvement, so no adverse effect occurred.
 
Manufacturer Narrative
Conclusion: the reported event was not confirmed.As an electro-mechanical instrument components can wear out or become damaged during use.The most likely cause is normal component wear out.Medtronic will continue to monitor for future trends and occurrences.Correction d4.2 (expiration date): the expiration date that was provided in this field in the initial report (2184009-2023-01008) was provided in error.This instrument is a repairable piece of capital equipment and it does not have an expiration date.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACT PLUS INSTRUMENT
Type of Device
TIMER, CLOT, AUTOMATED
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17994142
MDR Text Key326352807
Report Number2184009-2023-01008
Device Sequence Number1
Product Code GKN
UDI-Device Identifier00721902793066
UDI-Public00721902793066
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K940426
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/15/2016
Device Model NumberACT100
Device Catalogue NumberACT100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/27/2023
Initial Date FDA Received10/24/2023
Supplement Dates Manufacturer Received10/26/2023
Supplement Dates FDA Received10/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-