Model Number X SERIES
Device Problems Delayed Charge Time (2586); Unexpected Shutdown (4019)
Patient Problem Cardiac Arrest (1762)
Event Date 10/06/2023
Event Type  malfunction  
Event Description
Complainant alleged that while attempting to defibrillate a male patient (age unknown), the device delayed charging and inappropriately shut down.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Manufacturer Narrative
The device was returned to zoll medical united kingdom for evaluation.The device performed to specification.The logs indicate multiple low battery messages were prompted to the user during this event, including one right before a shutdown warning.The device passed all functional testing and bench handling using the returned battery.The customer has indicated the ambulance mount is working as expected and the device was capable of charging a battery as expected.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key17997087
MDR Text Key326411230
Report Number1220908-2023-04146
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P160022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/13/2023
Initial Date FDA Received10/24/2023
Supplement Dates Manufacturer Received10/13/2023
Supplement Dates FDA Received11/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-