Model Number X SERIES |
Device Problems
Delayed Charge Time (2586); Unexpected Shutdown (4019)
|
Patient Problem
Cardiac Arrest (1762)
|
Event Date 10/06/2023 |
Event Type
malfunction
|
Event Description
|
Complainant alleged that while attempting to defibrillate a male patient (age unknown), the device delayed charging and inappropriately shut down.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical united kingdom for evaluation.The device performed to specification.The logs indicate multiple low battery messages were prompted to the user during this event, including one right before a shutdown warning.The device passed all functional testing and bench handling using the returned battery.The customer has indicated the ambulance mount is working as expected and the device was capable of charging a battery as expected.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
|