Catalog Number 09015051190 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/08/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The cobas e 801 analytical unit serial number was (b)(6).Calibration was last performed on 10-jul-2023.Qc was acceptable.The sample was received for investigation.The sample was negative for both treponemal tests: elecsys syphilis: 0.056 coi (negative).Sekisui method: 0.0 t.U.The sample was negative with a non-treponemal test: sekisui method: 0.00 r.U.The investigation is ongoing.
|
|
Event Description
|
We received an allegation of a discrepant negative result for 1 patient sample tested for elecsys syphilis (syphilis) on a cobas e 801 analytical unit.The elecsys syphilis result was 0.0799 coi (negative).The inno-lia syphilis result was positive.
|
|
Manufacturer Narrative
|
The sample underwent further investigation.The result of the elecsys syphilis test was 0.090 coi (negative).The sample was diluted at 1:10 and 1:20 and tested with the elecsys syphilis and 2 other methods.All results were negative.Based on the investigation results, the sample is considered to be correctly negative with the elecsys syphilis assay.The assay performs within specification.The investigation did not identify a product problem.
|
|
Search Alerts/Recalls
|