Catalog Number 999800754 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Abscess (1690); Cyst(s) (1800); Inflammation (1932); Joint Laxity (4526); Metal Related Pathology (4530); Unspecified Tissue Injury (4559); Insufficient Information (4580)
|
Event Date 01/11/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint # (b)(4).D4-the device catalog number is unknown; therefore, udi is unavailable.E3 initial reporter occupation: lawyer.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Litigation received.Patient had a revision due to injury.Doi: unknown.Dor: (b)(6), 2023.Unknown hip.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary.No device associated with this report was received for examination.The asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot.Device history reviews for asr platform have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.
|
|
Event Description
|
Updated event description: patient has elevated ion lvl.Elevated esr.Suspicion for infected left tha w/ a psoas abscess.Snynovium was comprised of inflammatory grey tissue found on cup- spot welds.Mild bone loss.Wide capsulutomy to resect the inflammatory tissue and the hip was unstable.A cyst in the superior weight - bearing area was curetted free of soft tissue.Piriformis and gluteus maximus insertion were repaired.Doi: (b)(6) 2008.Dor: (b)(6) 2013.Left hip.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: d1, d2a, d4 (lot,catalog, expiration date, udi).
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : device history reviews for asr platform have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.
|
|
Search Alerts/Recalls
|
|