• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXPRESS LD VASCULAR; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXPRESS LD VASCULAR; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 20212
Device Problems Premature Activation (1484); Device Dislodged or Dislocated (2923)
Patient Problem Perforation of Vessels (2135)
Event Date 10/03/2023
Event Type  Injury  
Event Description
It was reported that stent dislodgement occurred, resulting in vessel trauma and requiring additional surgical intervention.The target lesion was located in the non-tortuous and mildly calcified common iliac artery.After local anesthesia was administered, an 8.0x40x135 cm express ld vascular was advanced to treat the lesion.However, the stent's proximal end prematurely opened up and the stent dislodged without the balloon being inflated.The dislodged stent caused intimal injury and an embolectomy was performed as the stent could not be snared.The dislodged stent was successfully removed and no further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS LD VASCULAR
Type of Device
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17999878
MDR Text Key326446204
Report Number2124215-2023-57454
Device Sequence Number1
Product Code FGE
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20212
Device Catalogue Number20212
Device Lot Number0031382226
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-