EDWARDS LIFESCIENCES SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
|
Back to Search Results |
|
Model Number 9750TFX26 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/27/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation is still ongoing.
|
|
Event Description
|
As reported by an edwards united kingdom affiliate, during a transfemoral tavr procedure with a 26mm sapien 3 ultra valve, resistance was felt when pushing the commander delivery system with crimped valve through the esheath, and upon exiting the esheath, it was visually noticed that the valve had a strut at the distal end that was protruding out at 90 degrees.However, it was decided to proceed.The valve was deployed successfully, and the strut straightened out with deployment.Upon removal of the esheath, a vertical rip up was noted in the esheath caused by the valve strut.There was no injury to the patient.The patient was doing well post procedure.Per report, the team felt that the calcium and tortuosity in the iliac caused resistance in the esheath and with the forces applied, the valve strut caught on calcium.
|
|
Manufacturer Narrative
|
Correction to h6 based on additional information.The device was not returned to edwards lifesciences for evaluation, however, a device history record (dhr) review was done and did not reveal any manufacturing nonconformance issues that would have contributed to the event.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of the failure mode is not required at this time.The reported event of frame damage was unable to be confirmed as no device/relevant imagery were provided.However, an existing technical summary written by edwards lifesciences captures the root cause analysis for complaints evaluated for resistance with delivery system and valve frame damage as a result from increased push force.The root causes identified in the technical summary were reviewed and the following were identified as applicable to this event: tortuous patient anatomy can create sub optimal angles that can lead to non coaxial alignment between the delivery system with crimped valve and sheath inner lumen during advancement, leading to resistance.As per description event, there was presence of tortuosity in the patient's vessels.Calcification can reduce the vessel lumen diameter and may increase restriction leading to resistance.Similar to tortuosity, calcification can also result in the creation of sub optimal angles during delivery system insertion that may lead to resistance.As per event description, there was presence of calcification in the patient's vessels.Excessive device manipulation or high push force can lead to the valve struts interacting with the sheath shaft and resulted the strut damage at the valve inflow side.It was reported that force was applied to overcome the resistance during delivery system advancement.The presence of the above factors can create challenging pathway during delivery system advancement, leading to resistance.More than one of these factors can compound to further exacerbate the patient procedural conditions and increase the likelihood of encountering resistance during delivery system advancement through the sheath resulted in frame damage.The technical summary also outlines the extensive manufacturing mitigations in'place to detect a defect or nonconformance associated with this issue.There are several 100% in process inspections (visual) performed in manufacturing process and product verification testing (functional and visual) on a sampling plan basis performed prior to lot release.These inspections and testing support that it is unlikely that a manufacturing non conformance contributed to the complaint.In addition, assessment of the detailed instructions for use (ifu), device preparation training manuals, and procedural use training manual revealed no identifiable deficiencies.These mitigations (from manufacturing and the ifu/training manual) as identified in the technical summary are still in place to mitigate this issue.As such, available information suggests that patient factors (calcification, tortuosity) and procedural factors (excessive device manipulation/high push force) may have contributed to the reported event.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.
|
|
Search Alerts/Recalls
|
|
|