• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH HEATER COOLER 3T (230 VOLT); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH HEATER COOLER 3T (230 VOLT); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problems Failure to Pump (1502); Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2023
Event Type  malfunction  
Manufacturer Narrative
A.1.-a.5.There was no known patient involvement.G.5.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in china.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland has received a report that a 3t heater cooler displayed error pump 1 uses too much power (e06) on patient tank during maintenance.There was no patient involvement.This unit was returned from factory after deepdisinfection in may and was installed as loaner unit in september.The distribution board and the stirrer motor were replaced and the problem was solved.
 
Manufacturer Narrative
All functional tests positively passed and the unit was put back into service.Complaints database analysis revealed no similar event since unit installation in 2021.Based on all the above facts and the results of a similar complaints database analysis, it cannot be ruled out that the most likely root cause of the reported event is a non-free to spin pump motor, likely due to environmental conditions, as water infiltration, humidity, condensation or high temperatures, that required a power overload to work, which could have resulted in an over-current ultimately caused damage to involved board.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER COOLER 3T (230 VOLT)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18002109
MDR Text Key326453437
Report Number9611109-2023-00564
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817901099
UDI-Public(01)04033817901099(11)211012
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/28/2023
Initial Date FDA Received10/25/2023
Supplement Dates Manufacturer Received12/22/2023
Supplement Dates FDA Received01/17/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/12/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-