ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number R SERIES
Device Problem Incorrect Interpretation of Signal (1543)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Complainant alleged that while attempting to defibrillate a patient (age & gender unknown), the device issued a "shock advised" prompt for a heart rhythm they believe was non-shockable.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Manufacturer Narrative
Zoll medical united kingdom evaluated the device and the customer's report was not confirmed.The device passed all testing successfully.The customer indicated that cpr may have been happening during the analysis of the patient, which can influence the device's algorithm.The r series operator's guide states, "do not use the unit in aed mode or advisory mode during patient movement.A patient must be motionless during ecg rhythm analysis.Do not touch the patient during analysis.If transporting the patient, cease all movement before beginning ecg analysis.Ecg rhythm analysis does not warn of patient asystole, which is not a shockable rhythm." the device was recertified and returned to the customer.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key18002729
MDR Text Key326470707
Report Number1220908-2023-04136
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P160022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/16/2023
Initial Date FDA Received10/25/2023
Supplement Dates Manufacturer Received10/16/2023
Supplement Dates FDA Received12/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-