Zoll medical united kingdom evaluated the device and the customer's report was not confirmed.The device passed all testing successfully.The customer indicated that cpr may have been happening during the analysis of the patient, which can influence the device's algorithm.The r series operator's guide states, "do not use the unit in aed mode or advisory mode during patient movement.A patient must be motionless during ecg rhythm analysis.Do not touch the patient during analysis.If transporting the patient, cease all movement before beginning ecg analysis.Ecg rhythm analysis does not warn of patient asystole, which is not a shockable rhythm." the device was recertified and returned to the customer.No trend is associated with reports of this type.
|