DATASCOPE CORP. - MAHWAH; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/16/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during a routine check, the cs300 intra-aortic balloon pump (iabp) unit had electrical test fail #50 coming.There was no patient involvement.
 
Event Description
N/a.
 
Manufacturer Narrative
A getinge field service engineer(fse) was dispatched to evaluate the iabp device.The fse found the machine shows electrical test failure # 50 with continuous beep.The fse reset the connections of the motor control pcb and main board.The fse checked afterwards, and found the problem remained the same.The fse tried reconnecting the motor control pcb and found the machine to be working ok.All tests passed.The unit was then handed over to the customer in good working condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18003841
MDR Text Key327311153
Report Number2249723-2023-04594
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/16/2023
Initial Date FDA Received10/25/2023
Supplement Dates Manufacturer Received02/01/2024
Supplement Dates FDA Received02/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/27/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-