Model Number 106 |
Device Problems
Premature Discharge of Battery (1057); Premature Elective Replacement Indicator (1483); Battery Problem: High Impedance (2947)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/02/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the patients generator was recently interrogated and showed that it lost a significant amount of battery life even thought it has only been recently implanted.Internal generator data was received and reviewed.No system diagnostic information was available.It was found that the patient's device is prematurely depleting.It was noted that on 10/02/2023 only 7.543% of the battery was consumed meaning 92.457% of the battery should be remaining, not 21%.On 10/02/2023, the pbcr using history was 92.4567176 and the pbcr using vbat was 21.47288.Per the equation, vbat = 3.518778 + (92.4567176)*0.44567 = 44.72396 versus 21.47288 measured ¿pbcr using vbat¿ on the decoder.Since the measured vbat on the decoder is lower than what the formula gives, the device is below the 95% prediction interval and this is evidence that the battery is depleting faster than expected.No other relevant information has been received to date.
|
|
Manufacturer Narrative
|
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanovas employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any defects¿ or malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
|
|
Event Description
|
Additional information was received noting that upon an updated interrogation of the device, the battery has rebounded to 75-100%.Diagnostics were noted to be within normal limits.The battery rebound event was confirmed to be related to a known event cause by the impedance behaviors in the beginning of the battery's life.Internal investigation identified that in some cases, pulse generators reach the 25% battery indicator earlier than expected, and later rebound to 75-100% without any significant programming changes.This behavior was determined to be due to an increased duration of high battery impedance during generator battery beginning-of-life.There is no evidence of a device malfunction.
|
|
Search Alerts/Recalls
|
|
|