• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS TOTAL PSA; PROSTATE-SPECIFIC ANTIGEN (PSA) FOR MANAGEMENT OF PROSTATE CANCERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS TOTAL PSA; PROSTATE-SPECIFIC ANTIGEN (PSA) FOR MANAGEMENT OF PROSTATE CANCERS Back to Search Results
Catalog Number 08791732190
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2023
Event Type  malfunction  
Manufacturer Narrative
The serial number of the e 801 analyzer was requested, but not provided.The investigation is ongoing.
 
Event Description
The initial reporter stated they received discrepant results for two patient samples tested with elecsys total psa on a cobas e 801 module.The clinician suspected that total psa patient results were incorrectly low.The customer noted that the ion blower on the analyzer lacked power, so they disposed of the reagent that was in use and started using the standby total psa reagent pack.The customer repeated 11 patient samples with the standby pack and of these, two had discrepant results.The first sample initially resulted in a total psa value of 4.88 ng/ml and it repeated as 0.02 ng/ml.The second sample initially resulted in a total psa value of 346.0 ng/ml and it repeated as 3.49 ng/ml.
 
Manufacturer Narrative
The questionable results were not reported outside of the laboratory to the patients.For sample 1, the initial total psa value was 0.02 ng/ml and the repeat value was 4.88 ng/ml.For sample 2, the initial total psa value was 3.49 ng/ml and the repeat value was 346.0 ng/ml.The customer stated there was absence of power to the ion blower on the analyzer.A field service engineer ran performance testing and precision studies.No problems were observed.The investigation is ongoing.
 
Manufacturer Narrative
For the last calibration performed on (b)(6) 2023, the calibration signals were lower than expected.Quality controls recovered well within range.A general reagent related issue or single reagent pack failure can be excluded as a root cause as controls were acceptable on the day of the event.No further issues occurred after the replacement of the reagent pack.The investigation could not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TOTAL PSA
Type of Device
PROSTATE-SPECIFIC ANTIGEN (PSA) FOR MANAGEMENT OF PROSTATE CANCERS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18012112
MDR Text Key326622739
Report Number1823260-2023-03435
Device Sequence Number1
Product Code LTJ
UDI-Device Identifier07613336165980
UDI-Public07613336165980
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
P990056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 01/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number08791732190
Device Lot Number652127
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/04/2023
Initial Date FDA Received10/26/2023
Supplement Dates Manufacturer Received10/27/2023
01/05/2024
Supplement Dates FDA Received11/20/2023
01/05/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-