Model Number 7122Q/58
Device Problems Product Quality Problem (1506); Retraction Problem (1536)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/20/2023
Event Type  malfunction  
Event Description
It was reported, the helix was already extended in the packaging and the helix was unable to be retracted back in.The lead was not used.There were no patient consequences.
 
Manufacturer Narrative
The reported events were helix was already extended in the packaging and the helix was unable to be retracted back in.As received, a complete lead was returned in one piece.The reported event of helix was already extended in the packaging was not confirmed.Visual inspection of the lead found the helix was stretched consistent with procedural damage and clogged with blood/tissue.The reported event of a helix mechanism issue was confirmed.X-ray examination found the helix was stretched consistent with procedural damage.The helix mechanism/ extension length test could not be performed due to the damaged helix, as received.The cause of the helix mechanism issue was isolated to blood/tissue in the helix region and stretched /damaged helix.Electrical testing did not find any indication of conductor fractures or internal shorts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18016642
MDR Text Key326705563
Report Number2017865-2023-50811
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734503174
UDI-Public05414734503174
Combination Product (y/n)Y
Reporter Country CodeFR
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7122Q/58
Device Lot NumberA000145042
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/26/2023
Supplement Dates Manufacturer Received10/27/2023
11/20/2023
Supplement Dates FDA Received10/30/2023
11/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-