It was reported that this patient reported to the hospital emergency room after receiving a shock.During the interrogation of the device, a code 1005 screen appeared on the programmer, indicative of a break in the energy or conductor system and revealed that this right ventricular (rv) lead exhibited high, out of range shock impedance (greater than 145 ohms).Increasing over the last 2 years.The patient was admitted to the hospital for monitoring.Lead integrity testing and device pocket maneuvers revealed noise.Subsequently this rv lead was surgically capped/abandoned (i.E., it remains implanted, but is no longer in service) and the device was explanted.Besides surgical intervention, no adverse patient effects were reported.This rv lead is not expected to be returned for analysis.
|