• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-S190-10
Device Problems Signal Artifact/Noise (1036); Break (1069); Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned as part of the asset return process due to deformation of video connector; water tightness is lost, adhesive on bending section cover has a chip, video connector case has a crack, due to wear of angle wire, bending angle in up direction does not meet specifications, venting connector of light guide connector is deformed.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Event Description
A user facility returned the olympus asset, endoeye flex deflectable videoscope, to the olympus service center.Upon inspection and testing of the returned device, it was observed that ¿due to damage on ccd unit, image noise occurs, and the image cannot be seen.¿ this report is being submitted for the event found during evaluation.There was no patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to correct h6: problem code (1036-signal artifact noise should not be selected on the initial medwch).A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be conclusively determined.It is likely the reported event occurred due to a damaged image sensor unit or broken mounting components (chip, capacitor, ect.) on the electrical board caused by stress from external factors or handling.The event can be detected by handling the device in accordance with the following instructions for use( ifu): "[inspection of the endoscope] confirm that the wli and nbi endoscopic images are normal.1.Before inspection, wipe the objective lens using clean lint-free cloths moistened with saline solution or sterilized water.2.Observe the palm of your hand in the wli and nbi endoscopic images.3.Confirm that light is output from the endoscope¿s distal end.4.Adjust the brightness level as appropriate.5.Confirm that the wli and nbi endoscopic images are free from noise, blur, fog, or other irregularities.6.Turn the angulation control levers slowly in each direction until it stops.7.Confirm that the wli and nbi endoscopic images do not momentarily disappear or display any other irregularities." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18018434
MDR Text Key326681930
Report Number9610595-2023-15967
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170310386
UDI-Public04953170310386
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/03/2023
Initial Date FDA Received10/27/2023
Supplement Dates Manufacturer Received11/14/2023
Supplement Dates FDA Received11/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-