Catalog Number 14687-15
Device Problems Fluid/Blood Leak (1250); Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2023
Event Type  malfunction  
Manufacturer Narrative
The device is not available for investigation.However, a photo of the affected product was provided.The evaluation is not yet complete.
 
Event Description
The incident involved a primary plum set, clave secondary port, clave y-site, secure lock, 103 inch.The reporter stated that the fluids could not be injected into the patient due to an upper section rupture of the tubing which caused patient discomfort or an uncontrolled condition and had to extend their hospital stay.There was patient involvement and unknown patient harm.
 
Manufacturer Narrative
A photo was returned by the customer showing the set in a plastic bag.No damage or anomalies noted.No samples were returned for investigation.The reported complaint of ruptured tubing on the (b)(6) was unable to be replicated or confirmed.Without the return of the used sample, a comprehensive failure investigation cannot be performed.The device history review for lot 13591617 was reviewed and no non conformities were found that would have led to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18021132
MDR Text Key326724191
Report Number9615050-2023-00576
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10887787005650
UDI-Public(01)10887787005650(17)260401(10)13591617
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K141789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number14687-15
Device Lot Number13591617
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/05/2023
Initial Date FDA Received10/27/2023
Supplement Dates Manufacturer Received11/17/2023
Supplement Dates FDA Received11/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-