Model Number LEV2512-MVE |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Paralysis (1997); Speech Disorder (4415); Ischemia Stroke (4418)
|
Event Date 04/01/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation, nor were any images of the device provided in place of a device return.Procedure and medical imaging was not provided for this investigation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): potential complications possible complications include but are not limited to the following: hematoma at the puncture site perforation or dissection of the vessel(s) intravascular spasm hemorrhaging rupture or perforation of aneurysm coil herniation device migration neurologic insufficiencies including stroke and death ischemia vascular occlusion vessel stenosis incomplete aneurysm occlusion pseudoaneurysm formation distal embolization headache infection reaction to contrast agents including severe allergic reactions and renal failure warnings should unusual resistance be felt at any time during access or removal, the introducer/guide catheter/microcatheter and lvis evo device should be removed as a single unit.Applying excessive force during delivery or retrieval of the lvis evo device can potentially result in loss or damage to the device and delivery components.It is imperative to use the lvis evo device with compatible microcatheters.If repeated friction is encountered during lvis evo device delivery, verify microcatheter is not kinked or in extremely tortuous anatomy.Confirm that the microcatheter does not ovalize.Confirm that there is adequate sterile flush solution.Do not reposition the lvis evo device in the parent vessel without fully retrieving the device.The lvis evo device must be retrieved into the microcatheter and re-deployed at the desired target location or removed completely from the patient.Precautions exercise caution when crossing the deployed/detached lvis evo device with adjunctive devices such as guidewires, catheters, microcatheters or balloon catheters to avoid disrupting the device geometry and device placement.Directions for use 15.Advance the delivery wire to transfer the lvis evo device from within the introducer into the microcatheter.Warning: do not torque the delivery wire while advancing or retracting the lvis evo device.A torque device should not be used.16.Continue advancing the delivery wire into the microcatheter until the proximal tip of the delivery wire enters the introducer.Loosen the rhv locking ring, remove the introducer, and set it aside.Warning: do not apply undue force.If resistance is encountered at any point during lvis evo device delivery or manipulation, withdraw the unit and select a new lvis evo device.18.Position the lvis evo device for deployment by aligning the lvis evo implant distal radiopaque end markers approximately 7 mm or adequate length past the aneurysm neck.Note: a proper push/pull technique, encompassing sufficient delivery wire push force, in addition to an opposing microcatheter withdrawal force, will facilitate properly deploying the lvis evo device to achieve full expansion and good vessel apposition.Note: slowly advancing the lvis evo device while adjusting the microcatheter position will ensure accurate deployment.Maintain simultaneous control of the lvis evo device and microcatheter in order to position and expand the device at the proper location.Caution: using a rapid microcatheter withdrawal technique to deploy the lvis evo device is not recommended and may result in device elongation.19.If lvis evo device positioning is not satisfactory, the lvis evo device may be recaptured and repositioned if it is not fully deployed.The lvis evo device may be recaptured until the point where the proximal end of the lvis evo device markers is aligned 3 mm proximally with the microcatheter distal marker band.Caution: if resistance is felt while recapturing the lvis evo device, do not continue to recapture the device.Withdraw the microcatheter slightly to unsheath the lvis evo device (without exceeding the recapture limit), and then attempt to recapture the lvis evo device.Caution: the lvis evo device must not be re-deployed more than three times.Note: the lvis evo device delivery wire should not be utilized as a guidewire.Do not torque the lvis evo device.A torque device should not be used.20.If lvis evo device positioning is satisfactory, carefully retract the microcatheter and advance the delivery wire together, to allow the lvis evo device to deploy across the neck of the aneurysm.Ensure the device proximal radiopaque end markers are approximately 7 mm or adequate length proximal to the aneurysm neck to ensure an adequate landing zone.The lvis evo device will expand and total length may foreshorten up to 60% from its undeployed length (refer to table 1) as it exits the microcatheter.Ensure microcatheter is retracted and clear from the proximal flared ends.Note: visualize and refer to the implant radiopaque end markers to maintain adequate implant length, approximately 7 mm or adequate length on each side of the aneurysm neck or target location to ensure appropriate neck coverage.Warning: do not detach the lvis evo device if it is not properly positioned in the parent vessel.Observe the delivery wire distal tip to assure it remains within the desired location of the parent vessel.21.Prior to removing the delivery wire and if necessary, carefully position the microcatheter distal to the lvis evo device to maintain access through the lvis evo device.Remove and discard the delivery wire.Warning: the lvis evo device delivery wire should not be utilized as a guidewire.Do not torque the lvis evo device.A torque device should not be used.23.Use the guidewire and microcatheter to access the aneurysm through the lvis evo device cells.Warning: observe lvis evo device marker position during placement of the microcatheter into the aneurysm to ensure that the lvis evo device does not migrate or dislodge from its deployed position.Note: access to the aneurysm may be facilitated by the use of a microcatheter that has been shaped.25.Warning: observe lvis evo device marker position during the coiling procedure to ensure that the device does not migrate from its deployed position.After placing the last coil, verify that the lvis evo device has remained patent and properly positioned.Advance a guidewire, if necessary, to the microcatheter tip and carefully remove the microcatheter.Note: a microcatheter may be positioned into the aneurysm sac prior to delivery of the lvis evo device.The microcatheter will be supported by the lvis evo device during delivery of embolic coiling.After completing the coiling, the coiling microcatheter should be carefully removed to avoid dislodging the lvis evo device.27.Caution: carefully watch the lvis evo device distal and proximal markers when passing through the deployed lvis evo device with embolic coiling microcatheters to avoid displacing the lvis evo device.Investigation conclusion: the physical device was not available for evaluation to determine if a condition existed that would have caused or contributed to event.Supplemental imaging was also unavailable for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.This information may be updated if additional information is provided at a later date.
|
|
Event Description
|
As reported through the sealant clinical study, the patient was treated for a 2.6 x 2.5 x 2 saccular bifurcation aneurysm, located on the left middle cerebral artery (mca).The aneurysm never ruptured and not previously treated.Pre-procedure neurological evaluation was: mrs score 0 - nihss score 0.The treatment consisted of using a lvis evo stent and hydrosoft helical coils during the procedure under stenting post-coiling technique.The parent artery, 2.3mm distally and 2.5mm proximally, at the end of the procedure remains without stenosis.Patient discharged from the hospital on (b)(6) 2022, with a neurological evaluation; mrs 0, nihss score 0.Reportedly, on (b)(6) 2023, 5 months post procedure, the patient experienced neurological event (ischemic stroke), left side hemiparesis, speech difficulty and lack of concentration.Symptoms occurred 15-20 times (duration about 10 minutes) following the cessation of clopidogrel after planned discontinuation 6 months ago.The adverse event was stated to be possibly related to the lvis evo device and not related to the hydrosoft coils study device.Digital subtraction angiography of the vessels supplying the brain showed the implanted stent is free flowing.Minimal intimal hyperplasia is seen in the stent implanted in the left middle cerebral artery.No hemodynamic relevance.The pretreated aneurysms are completely excluded from the blood circulation.Monotherapy with asa 100mg should be continued permanently.Reportedly, the outcome is resolved without sequelae.
|
|
Manufacturer Narrative
|
A medical report was received and has been reviewed with following summary: a detailed procedure medical information review has been performed.Procedure data review indicates that the patient was treated for a 2.6 x 2.5 saccular bifurcation aneurysm located within the middle cerebral artery as part of the selant clinical study.The patient treatment consisted of using a lvis evo stent and hydrosoft coils performed on operative date (b)(6) 2022.Medical review of procedure data indicates the patient received stent-assisted coiling of a right medial aneurysm and right carotid t aneurysm.On (b)(6) 2023, five months post-operative, the patient experienced ischemic stroke with left side hemiparesis.Digital subtraction angiography of vessels supplying the brain was established with standard and targeted magnification techniques of the right internal carotid artery and the internal carotid artery left established post-operative device visualization.Intraoperative device assessment utilizing (mri techniques: axial t2 flair blade fs, axial swi, sagittal t2 tse, axial dwi with adc map, tof) further indicated, diagnostic minimal intimal hyperplasia is seen in the stent implanted in the left middle cerebral artery and no hemodynamic relevance.Intraoperative assessment of other vessels segments are reported as unremarkable according to course, caliber and hemodynamics as described within the procedure data presented.Based on the medical review of procedure data and in my clinical opinion, the performance of the post-operative, procedure assessment of the implanted devices demonstrated that the outcome of the coil embolized aneurysms continue to appear completely eliminated from the blood circulation and the implanted stents are free flowing.Medical review of procedure data does not report any occurrence of a device malfunction and the pretreated aneurysms are reported as completely excluded from the blood circulation.
|
|
Search Alerts/Recalls
|
|
|