Model Number 2088TC/65
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2023
Event Type  malfunction  
Event Description
It was reported that during initial implant procedure twisted insulation was noted on patient's right ventricular (rv) lead.The lead was not used and the procedure was completed using an alternate lead on (b)(6) 2023.There were no patient consequences.
 
Manufacturer Narrative
The reported event of insulation twisted was confirmed.As received, a complete lead was returned for analysis.Visual inspection of the lead found outer insulation was twisted throughout the entire lead.Electrical testing did not find any indication of conductor fractures or internal shorts.The cause of reported event was due to procedure damage.No other anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18025812
MDR Text Key327246629
Report Number2017865-2023-50917
Device Sequence Number1
Product Code NVN
UDI-Device Identifier05414734505055
UDI-Public05414734505055
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2088TC/65
Device Lot NumberP000176914
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/12/2023
Initial Date FDA Received10/27/2023
Supplement Dates Manufacturer Received12/13/2023
Supplement Dates FDA Received12/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexMale
-
-