• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC CARTO® 3 SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC CARTO® 3 SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Catalog Number FG540000K
Device Problem Pacing Problem (1439)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/02/2023
Event Type  malfunction  
Event Description
It was reported that a patient underwent a paroxysmal supraventricular tachycardia cardiac ablation procedure with a carto® 3 system and immediately after the procedure, a pacing spike entered after 2 seconds from pacing through carto3, and cardiac cavity a wave was captured.The procedure was continued and completed without any problems.No adverse patient consequence was reported.
 
Manufacturer Narrative
E 1.Initial reporter phone: (b)(6).The hardware investigation has begun but it has not been completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
It was reported that a patient underwent a paroxysmal supraventricular tachycardia cardiac ablation procedure with a carto® 3 system and immediately after the procedure, a pacing spike entered after 2 seconds from pacing through carto3, and cardiac cavity a wave was captured.The procedure was continued and completed without any problems.No adverse patient consequence was reported.Hardware evaluation details: pacing was performed at the facility using jumper cables without going through carto 3 system.The customer has informed that the system is operating normally.In addition, an investigation was initiated by the manufacturer to look into the issue further.The data was requested for investigation; however, no data was available for review.As a result, it was not possible to reproduce or analyze the issue.The root cause of the reported issue was not determined.The system is operational.The history of customer complaints reported during the last year associated with carto 3 system # 55575 was reviewed.No similar complaints were found.A manufacturing record evaluation was performed on carto 3 system #55575, and no internal actions related to the reported complaint condition were identified.Additionally, the manufactured date has been provided.Therefore, field h4.Device manufacture date has been populated.The complaint investigation results will be used for monitoring and detecting statistical signals per complaint trending and signal detection process.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTO® 3 SYSTEM
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER (ISRAEL) LTD.
4 hatnufah street
yokneam 20667 17
IS   2066717
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18026191
MDR Text Key327049464
Report Number2029046-2023-02467
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K042999
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFG540000K
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/02/2023
Initial Date FDA Received10/27/2023
Supplement Dates Manufacturer Received02/05/2024
Supplement Dates FDA Received02/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2018
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-