• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer¿s allegation was confirmed.During the evaluation it was found that the connecting tube has coating peeling, (maximum width 4mm or more).Additional findings were as follows: due to a pinhole on channel-tube, water tightness is lost, due to a pinhole on connecting tube, water tightness is lost, the adhesive on bending section cover has a chip, the connecting tube has a wrinkle, the video connector case has discoloration, the universal cord has a dent, due to wear of angle wire, bending angle in up direction does not meet the standard value, and an abnormal sound is made due to damage on angulation lever.Based on the results of the investigation, the root cause or factors that caused coating of soft insertion tube peeled off (long diameter 4mm or more) could not be identified.Therefore, a definitive root cause cannot be identified.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Olympus will continue to monitor the field performance of this device.
 
Event Description
The customer reported to olympus that the uretero-reno videoscope had a leak.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was found: the insertion site soft tube coating peeled off.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18026469
MDR Text Key326818102
Report Number3002808148-2023-11956
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170340321
UDI-Public04953170340321
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K072957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/11/2023
Initial Date FDA Received10/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-