Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were reviewed.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.The device interrogation revealed the eos battery status, detected on october 16, 2023.The device was implanted for 46 months and 8 charging cycles were recorded in the devices memory.Next, the current consumption of the icd was analyzed and found to be normal and as expected.However, an inconsistency between current consumption and battery voltage was noted.Therefore, the icd was opened and the inner assembly was inspected.The visual inspection of the inner assembly showed no anomalies.The measurement of the battery voltage showed a depleted battery.The current consumption of the electronic module was verified by direct measurement and proved to be as expected.There was no indication of a malfunction of the electronic module.The battery was sent to the manufacturer for further analysis.The manufacturing records of the battery were inspected, documenting that the battery parameters were within specification during the battery manufacturing.No anomalies were documented during the production process, associated with this battery.The visual inspection of the battery did not reveal any external signs of damage.The voltage measurement confirmed the battery depletion.At a next step, the battery was opened for destructive analysis.The inspection of the inner assembly identified damaged insulation, which led to an increased internal self-depletion within the battery and therefore contributed to the clinical observation.In conclusion, the analysis revealed an elevated internal self-depletion within the battery to be the root cause of the clinical observation.
|