Model Number CD2357-40C
Device Problems Incorrect Interpretation of Signal (1543); Inappropriate/Inadequate Shock/Stimulation (1574); Delayed Charge Time (2586)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2023
Event Type  Injury  
Event Description
Related manufacturer reference number: 2017865-2023-51119.It was reported that patient presented to emergency room after experiencing in appropriate shock from implantable cardioverter defibrillator (icd).Upon interrogation, it was found that the icd failed capacitor maintenance due to long charge time.It was also noted that patient's atrial lead exhibited high threshold.Programming changes were made.The patient was stable.
 
Event Description
New information received notes that patient received inappropriate shock due to incorrect interpretation of signal from patient's implantable cardioverter defibrillator (icd).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA DR ICD, US
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18031157
MDR Text Key326867389
Report Number2017865-2023-51118
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734508063
UDI-Public05414734508063
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2015
Device Model NumberCD2357-40C
Device Lot Number4284057
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/13/2023
Initial Date FDA Received10/30/2023
Supplement Dates Manufacturer Received11/03/2023
Supplement Dates FDA Received11/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/08/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age76 YR
Patient SexMale
-
-