Model Number X SERIES |
Device Problem
Electrical Shorting (2926)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during functional testing, the device displayed a "short detected" message.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical canada; the customer's report was duplicated and attributed to a loose multifunction cable connector.The multifunction cable connector was replaced to resolve the report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|