Model Number 861288 |
Device Problems
Failure to Deliver Energy (1211); Failure of Device to Self-Test (2937)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/18/2023 |
Event Type
malfunction
|
Event Description
|
It was reported the device test discharges incorrect values.There was no patient involvement.
|
|
Event Description
|
It was reported the device test discharges incorrect values.There was no patient involvement.
|
|
Manufacturer Narrative
|
This report is based on information provided by philips remote service engineer (fse) and site technician has been investigated by the philips complaint handling team.The site technician evaluated the device on site.They found the high voltage capacitor issue with low impedance values.Based on the information available and the testing conducted, the cause of the reported problem was the high voltage capacitor.The reported problem was not confirmed.The device could not be confirmed to have this issue.The device is out of service since it needs a battery (addressed in pr (b)(4).The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
|