In speaking with olympus technical assistance center (tac), the customer requested an eta for service because "both preprocessors were down" and they needed "to know if they need to cancel cases for tomorrow".That same day, an olympus field service engineer (fse) had a videoconference call with the customer's gastrointestinal technicians (gi tech) for device troubleshooting.The fse did not find evidence of deformity during examination of the cassette drawer or other components.The fse also found no components that were broken or detached.The gi techs were led through the load process with a new set of acecide-c bottles.The load process was completed as directed.The tray operation was smooth and no error messages occurred.The reprocessor was used for a full reprocessing cycle which was successfully completed.During the videoconference, no device problems were identified and the device was found functional.The reported issue was resolved with customer, therefore the device was not returned to olympus for the evaluation of the reported event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over two years since the subject device was manufactured.The investigation determined that the acecide-c bottle packaging was likely transported and/or stored on its side rather than upright, which caused inflation or bulging of the bottle.The acecide-c bottle has a vented cap that relieves pressure from the acid solution; if the vented cap is covered, this may cause bulging or leakage.However, the root cause of the event could not be specified conclusively.Olympus will continue to monitor field performance for this device.
|