SHIRAKAWA OLYMPUS CO., LTD. ULTRASONIC PROBE Back to Search Results
Model Number UM-S20-17S
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2023
Event Type  malfunction  
Manufacturer Narrative
The device is not expected to be returned for evaluation.On 12oct2023 an (ess) endoscopic support specialist was dispatched to the site to perform a radial probe reprocessing and infection control in-service for the staff.They observed the manual cleaning and preparing of the probe for ethylene oxide (eto) sterilization processing.It was found that the customer uses a non-olympus eto sterilization company to sterilize their radial probes.Recommended that the customer contact the manufacturer of that equipment for proper use.The investigation is ongoing and follow-up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus that they thought oil was leaking from the ultrasonic probes due to oil stains found on the envelopes from sterilization processing.This issue was found during reprocessing.There were no reports of patient harm.Related patient identifiers: (b)(6).
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause was unable to be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18034799
MDR Text Key326962054
Report Number3002808148-2023-11997
Device Sequence Number1
Product Code ITX
UDI-Device Identifier04953170368479
UDI-Public04953170368479
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K982323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/05/2023
Initial Date FDA Received10/30/2023
Supplement Dates Manufacturer Received11/16/2023
Supplement Dates FDA Received11/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/16/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-