ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Catalog Number 17AGFN-756
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2023
Event Type  Injury  
Event Description
It was reported that on (b)(6) 2023, a 17mm sjm regent heart valve was implanted in a patient.Prior to implant, the physician rotate the valve with the valve rotator while it was in close position and no resistance was felt.Post implant, the valve leaflet broke without a specific reason into 3 pieces and was recover from the patient.The device was explanted and replaced with a new 17mm sjm regent heart valve that completed the procedure successfully.The patient is stable.
 
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
An event of leaflet fracture during procedure was reported.The investigation at abbott found that one leaflet had been dislodged from the orifice and was not returned for analysis.The intact leaflet was properly inserted into the orifice and able to fully open and close with no resistance noted.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the leaflet dislodgment could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18038765
MDR Text Key326963557
Report Number2135147-2023-04783
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number17AGFN-756
Device Lot Number8132957
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/07/2023
Initial Date FDA Received10/31/2023
Supplement Dates Manufacturer Received11/07/2023
Supplement Dates FDA Received11/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/10/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-