The device was returned to the manufacturer in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, a failure occurred on functional testing.The device's main board was replaced to address the issue.Damage was seen to the top enclosure and ui panel.Therefore, the top enclosure and ui panel were replaced.The device was visually inspected and confirmed that there was no peeling or deterioration of the foam.Replaced rp-airpath kit, a30, a40, v30, ola+ p/n 1152388 per lbl1155427.Performed functional test of the device per bipap a30 & bipap a40 original and silver series manual, and the device passed it successfully.Software version was upgraded to 3.6.5.
|